From: (Saudi Arabia)   Date: 28 January 2005, 04:41 [GMT]
Subject: Biological indicator

to 3m or concerm

my question is can we use the 1292 to 121 temp prevacuum cycle, on manufacturer instruction say not to you use on this temp.
instend 1292E. but the representative here in kingdom says
ignore the manufacturer instruction because they have same amount of spore. pls somebody explain this..

thanks

From: Miki   Date: 5 February 2005, 21:35 [GMT]
Subject: Re: Biological indicator

The Manufacturer Declaration is the only valid document. A mistake may happen, but also in this case, only the manufacturer can state it.