From: (United States)   Date: 20 February 2006, 22:21 [GMT]
Subject: Sterrad

What are the specifications for sterilizing cameras, scopes and light cords in the sterrad 50. We have the Strker High Definition System.

From: (United States)   Date: 8 March 2006, 14:04 [GMT]
Subject: Re: Sterrad

Barbara,
The first step is to check the materials compatibility of the Stryker products that you wish to sterilize. The ASP Customer Care Center can assist you to determine if your particular models of cameras and endoscopes can be safely sterilized in STERRAD. You will need to provide them with the manufacturer's name and the model numbers for each item you wish to sterilize. You can contact the CCC at 1-888-788-7723.
The stainless steel lumen sizes that are approved for sterilization in the STERRAD 50 are: inside diameter 1 mm or larger, length 125 mm or shorter; inside diameter 2 mm or larger, length 250 mm or shorter; inside diameter 3 mm or larger, length 400 mm or shorter.

Please use the Follow Up form if you want to post a reply or answer to question Q01120. For new questions please use the Submit New Question form.