WFHSS - World Forum for Hospital Sterile Supply

Members
Communications [30/09/2008] Open letter to heads of sterilization departments in countries where there are no national societies
Conferences & Congresses [01/10/2008] Conference Info Update: IASSM Conference 2008 - "Audit aftershock" - 7/8 November 2008 in Dublin-Blanchardstown, Ireland
Courses & Seminars
Questions & Answers [07/10/2008] new question 1946: Statim and eye instruments
Discussion Forum
CEN
Publications
Education
WFHSS WebShop
Industrial Partners
Commercial Announcements
Related Sites & Links
Vacancies
Humour [03/10/2008] new cartoon: Operating Manual
Industrial Partners:

7766912

Developed by Baumeister Mediasoft Engineering

WFHSS - World Forum for Hospital Sterile Supply

Central Service - Issue 6/99 - English VersionZentralsterilisation - Ausgabe 6/99 - Deutsche Version

HomeLogosNewsSearchSitemapContact

WFHSS - World Forum for Hospital Sterile Supply:: Publications :: Periodicals :: Central Service :: Central Service Archive :: Central Service 6/99
Central Service - Issue 6/99

Central Service - Issue 6/99

NEWS UPDATE

What's New in Standardisation: prEN ISO 15883 - Washer-Disinfectors
New NHS Documents
New Societies for New Challenges
FDA and IAHCSMM on Reprocessing of Single-Use Devices
The CSC - 30 Years on the Move
ISSM Conference 1999 in Blackpool, England
Sterile Supply: Challenges in the New Millenium (in German only)
Perspectives (in German only)

ORIGINAL ARTICLES

M.J. Alfa
Process Validation for Reprocessing of Patient-Used Cannulatomesn
Zentr Steril 1999; 7 (6): 368-382

Abstract
Fiscal pressures have stimulated many centres in North America to reprocess complex, difficult to clean medical devices such as papillotomy knives (cannulatomes), despite the fact that these are sold as single-use. The objectives of this study were to assess the current reprocessing methods to determine if they were effective, and to assess alternative approaches to determine if they could be validated as reliable for reprocessing patient-used cannulatomes.
The current in-house cleaning/ethylene oxide sterilisation methods were inadequate, as none of the nine routinely processed cannulatomes tested from two different centres were sterile. Three new reprocessing methods were evaluated, which all incorporated a pump to facilitate fluid flow through the narrow lumen of the cannulatomes. The most stringent method incorporated a pre-rinse of the cannulatome with 10 ml distilled water, exposure to detergent for 15 minutes at room temperature using a pump to circulate the detergent, followed by pump circulated rinsing with distilled water at room temperature for 10 minutes, followed by an alcohol rinse, forced air drying and two hours exposure to ethylene oxide.
Simulated-use testing was performed utilising an artificial test soil containing Enterococcus faecalis and Bacillus subtilis spores to provide an organic/inorganic challenge that was similar to what would be expected from patient-use. It demonstrated that the most stringent reprocessing method could eradicate 106 cfu/device of B. subtilis, however, 3/5 of the devices were not sterile as < 10 cfu/device of other organisms were detected. The same issue of low level residual organisms was found for patient-used cannulatomes that were reprocessed by the most stringent method.
In conclusion the study indicates that even with the most stringent method tested there was difficulty assuring sterility 100% of the time. Centres reprocessing single-use cannulatomes should evaluate whether their reprocessing methods are adequate.

Key Words: cannulatome, sterilisation, reprocessing, single-use, process validation, test soil

FROM THE FIELD

S.M. Kovach
Parametric Release from a Hospital Perspective
Zentr Steril 1999; 7 (6): 383-392

Abstract
Parametric release of steam sterilised items provides hospitals with certain advantages when compared to sterility release procedures currently used in the United States. However, the requirements to implement a quality system leading to parametric release will not be easily met by most hospitals today without major changes taking place. To implement an enhanced quality system leading to parametric release, hospitals in the United States are going to have to take the time and make the effort to understand the principles of parametric release, see its advantages and commit themselves to the changes it will require. The paper deals with the authors view on what are some of the obstacles that a healthcare facility in the United States could face in trying to get to a parametric release type system in their institution.

Key Words: Sterilisation, parametric release, validation, commissioning, performance qualification

DISCUSSION

U. Kaiser
Simple Method to Assess Efficacy of Sterilisation Processes for Hollow Instruments
Zentr Steril 1999; 7 (6): 393-395

Abstract
The penetration of the sterilant into hollow devices may be insufficient and can compromise the success of sterilisation depending on the process selected. The use of various process challenge devices (PCD) with different penetration characteristics offers the opportunity to evaluate the possibilities and limitations of steam-, ETO- H2O-, Formaldehyde- H2O- and H2O2-H2O sterilisation processes. The tests may be carried out with suitable biological and/or chemical indicators. Hints are given at how to compare PCD with real surgical instruments.

Key Words: Steam sterilisation, hollow device, process challenge device, PCD, air removal, non-condensable gases

EDUCATION

P. Heeg
Series on Infectiology - Infections by Facultatively Pathogenic Enterobacteria

INDUSTRY IN FOCUS

Industry in Focus
Uncomplicated Humanitarian Aid
Instrument Information System
Non-Leaded Bowie & Dick Test
Cleaning Brushes - The Most Innovative Product at the Basel Congress
Exact Pollutant Suctioning with the WIBOject System
ILS20 ISO-Compliant Test and Validation Steriliser Systems

* Denotes the author to whom questions and reprint requests should be addressed.


Publisher
Information on this page provided with kind permission of: mhp-verlag GmbH
Address: Marktplatz 13
65183 Wiesbaden
Germany
Web: www.mhp-verlag.de
Email:
mhp stands for publications in medicine, hygiene and prevention. For more than 20 years, the mhp publishing house has devoted itself to supporting developments aiming for a better standard of hygiene and infection control in hospitals and other medical facilities.
Archive
WFHSS - World Forum for Hospital Sterile Supply:: Publications :: Periodicals :: Central Service :: Central Service Archive :: Central Service 6/99

© 2008 WFHSS • Updated: 03 July 2007, 13:35 [GMT]