WFHSS - World Forum for Hospital Sterile Supply

Members
Communications [30/09/2008] Open letter to heads of sterilization departments in countries where there are no national societies
Conferences & Congresses [01/10/2008] Conference Info Update: IASSM Conference 2008 - "Audit aftershock" - 7/8 November 2008 in Dublin-Blanchardstown, Ireland
Courses & Seminars
Questions & Answers [07/10/2008] new question 1946: Statim and eye instruments
Discussion Forum
CEN
Publications
Education
WFHSS WebShop
Industrial Partners
Commercial Announcements
Related Sites & Links
Vacancies
Humour [03/10/2008] new cartoon: Operating Manual
Industrial Partners:

7766878

Developed by Baumeister Mediasoft Engineering

WFHSS - World Forum for Hospital Sterile Supply

Central Service - Issue 2/01 - English VersionZentralsterilisation - Ausgabe 2/01 - Deutsche Version

HomeLogosNewsSearchSitemapContact

WFHSS - World Forum for Hospital Sterile Supply:: Publications :: Periodicals :: Central Service :: Central Service Archive :: Central Service 2/01
Central Service - Issue 2/01

Central Service - Issue 2/01

EDITORIAL

World, European and regional congresses

NEWS UPDATE

What's New in Standardisation: Creutzfeldt-Jakob Disease
Recent developments in the UK
BVMed demands that reprocessors fulfil the same requirements as manufacturers of medical devices
New Measure Taken by the British Government on Use of Single-Use Instruments
Reprocessing Medical Devices - Legal Framework
Successful Start to IFSS. 1st Congress in Petaling Jaya, Malaysia; March 25 - 29, 2001

MAIN ARTICLES

Y. Uetera*, H. Saito, M. Ookuni, Ch. Tanaka:
Evaluation of thermal disinfection procedure in washer disinfectors using a wireless thermologger
Zentr Steril 2001; 9 (2): 88-99

Abstract
Accurate thermometry is essential in the thermal evaluation of washer disinfectors. The wireless thermologger is a useful tool for the thermal evaluation of washer disinfectors because it is independent of the thermoregulatory system of the washer disinfector. The present study focuses on when washer disinfectors should be thermally evaluated, using a wireless thermologger, to assure optimal management of thermal disinfection procedures. Method: The actual disinfection temperature and thermal exposure time were evaluated using the wireless thermologger in one multi-chamber (HAMO T-840cs) and two single-chamber (HAMO T-21's) washer disinfectors installed in Tokyo University Hospital. Disinfection temperature and thermal exposure time were programmed at 93 °C for either 5 or 10 minutes in these three washer disinfectors. Thermal evaluation of the washer disinfectors was sequentially conducted three times using the wireless thermologger system. The wireless thermologger was placed in a wirebasket, which was then loaded into the washing rack. The actual disinfection temperature and thermal exposure time measured by the wireless thermologger were compared with the programmed disinfection temperature and thermal exposure time.
Results: In all three washer disinfectors, the actual disinfection temperature failed to satisfy the programmed value in the first thermal evaluations. Sequential thermal evaluations of washer disinfectors with a wireless thermologger revealed that actual disinfection temperature also failed to meet the programmed value when a washer disinfector was installed or the temperature control system was manipulated.
Conclusion: The results of the present study suggest that thermal disinfection procedures for washer disinfectors must be evaluated, using a wireless thermologger system periodically as well as at the time of installation and whenever the temperature control system is manipulated or a control panel is replaced.

V. Chartier, D. Verjat, C. Fargeot and J.C. Darbord*:
Experimental Results of Detergent and Pre-Disinfection Treatments Before Sterilisation of Medical Devices
(Zentr Steril 2001; 9 (2): 100-107)

Abstract
The presence of biological soils on instrument surfaces is the main cause of contamination related to inadequate disinfection or sterilisation. Glutaraldehyde is the most widely used high level disinfectant in France. It is known to stick on prion proteins and to reduce treatment efficacy. We have compared the efficiencies of detergents, pre-treatment products and disinfectants, in particular glutaraldehyde. The intent was to define the rank of each in the treatment cycle of reusable medical devices and to validate "in use" parameters. Microplates were contaminated by yeast extract, bovine albumin supplemented with sheep blood, fibrin and bovine serum, and exposed to products (6 detergents, 2 chemical inactivators of prions and 3 disinfectants) at different concentrations and for different times. The overall efficacy of treatments was compared to untreated control microplates. The products which are recognised as chemical prion inactivator are able to eliminate all types of soils. In particular, sodium hypochlorite was found to be interesting, including at concentrations lower than 2 % (v/v), recommended by W.H.O. These observations could broaden the use of chlorine by reduction of its corrosive effects on medical instruments.

K. Sherlock:
Event Related Sterility Testing
(Zentr Steril 2001; 9 (2): 108-116)

Abstract
Event related sterility (ERS) is based on the concept that events, not time are responsible for contamination of sterilised supplies. In moving from a expiration dating policy to an ERS policy it is recommended that sterility testing of sample packs be carried out as part of the validation process. This is to ensure that packs which have been processed, wrapped, handled and stored in an appropriate manner will remain sterile over an extended period of time. A test method was developed and testing was carried out on a range of sample packs for four hospitals in the Brisbane area. In total 179 packs were tested with 4 positives giving a contamination rate of 2.2 %.

MAILBOX

Drying time

EDUCATION

P. Heeg
Infections by Candida species

RECOMMENDATIONS BY THE QUALITY TASK GROUP (AK »QUALITÄT«)

Quality Assurance on the Unclean Side of a CSSD

PRESS REVIEW

BSE, angiographic catheters, pressure-Volume catheter, resterilisation of single-use devices


Publisher
Information on this page provided with kind permission of: mhp-verlag GmbH
Address: Marktplatz 13
65183 Wiesbaden
Germany
Web: www.mhp-verlag.de
Email:
mhp stands for publications in medicine, hygiene and prevention. For more than 20 years, the mhp publishing house has devoted itself to supporting developments aiming for a better standard of hygiene and infection control in hospitals and other medical facilities.
Archive
WFHSS - World Forum for Hospital Sterile Supply:: Publications :: Periodicals :: Central Service :: Central Service Archive :: Central Service 2/01

© 2008 WFHSS • Updated: 29 October 2006, 04:09 [GMT]