Central Service - Issue 5/01 - English VersionZentralsterilisation - Ausgabe 5/01 - Deutsche Version

HomeLogosNewsSearchSitemapContact

WFHSS - World Forum for Hospital Sterile Supply:: Publications :: Periodicals :: Central Service :: Central Service Archive :: Central Service 5/01
Central Service - Issue 5/01

Central Service - Issue 5/01

EDITORIAL

Conference of the German Society for Sterile Supply (DGSV)

NEWS UPDATE

What's New in Standardisation: Process Challenge Devices
Instrument Processing Forum Moves to Berlin
Working Group "Hospital Hygiene" of AWMF
Instrument Processing Working Group Online
Sterile Supply Faces New Challenges: DGSV Conference in Tübingen, 3 - 5 October 2001

MAIN ARTICLES

B. Kirk:
Non-Condensable Gases (NCG) in the Steam Supply of a Sterilizer. Development and Validation of an Introduction Method - Response of the Bowie & Dick Textile Test Pack
(Zentr Steril 2001; 9 (5): 317-335)

Abstract
The nature, source and consequences of the presence of Non Condensable Gases (NCG's) in the steam supply of a porous load steam sterilizer are discussed. Two methods for the introduction of NCG's into a sterilizer steam supply are described. Two methods for monitoring the level of NCG in a sterilizer steam supply are evaluated and the relationship between them outlined. The consequences of underestimating the levels of NCG by the method prescribed in European Standard EN 285 and the advantages of using an in line instrument providing continuous measurement of levels are discussed. The effect of increasing levels of NCG in steam on the temperature profile within a Bowie and Dick Test (BDT) textile pack when tested in four different types of sterilizer cycle and when using two methods of introduction, is described. The results indicated that the level of NCG measured in the steam supply is proportional to the rate of steam consumption. The results also indicate that the temperature depression created within the centre of a BDT textile pack is proportional to the level of NCG measured in the steam and the type of sterilizer cycle employed.

G. Hüker, R. Machmerth, A. Kramer:
Validation of Sterilisation Processes with Ethylene Oxide according to DIN EN 550 in Healthcare Establishments
(Zentr Steril 2001; 9 (5): 335-345)

Abstract
Based on the example of a sterilisation process with ethylene oxide in a hospital which was validated according to DIN EN 550, it was revealed that general evaluation of the process is not possible using only microbiological tests. While the microbiological results are suggestive of impeccable sterilisation, considerable deviations from the specifications of the standard were noted on measuring the temperature, pressure and humidity at different locations in the steriliser chamber and in the items being sterilised. As regards the parameter that is most important for microbial kill, i.e. the ethylene oxide concentration, it was demonstrated on using this highly precise measuring method for the first time that the concentration was markedly exceeded at different locations within the chamber.
Described here are the measuring method and the measurement accuracy achieved with it, as well as the procedure adopted for validation of sterilisation processes with ethylene oxide, including the results obtained for the respective example.
The importance of online measurements for validation and parametric release of sterilised medical devices is underlined. Using this measuring system, it has been possible for the first time to meet the general requirements for characterisation of a sterilant as per ISO/FDIS 14937, issue 2000:9 (21) in respect of sterilisation with ethylene oxide.

FROM THE FIELD

H. Pahlke, Th. W. Fengler:
Verification of BD Test in the CSSD
(Zentr Steril 2001; 9 (5): 347-352)

Abstract
This paper presents a novel method of multi-dimensional representation of the parameters pressure, temperature and time. A steam penetration test (Bowie-Dick test) provides no information on the time course. Using a BD datalogger (e.g. ebro Electronic, Ingolstadt, Germany) these parameters of the BD test can be documented; inert gases cannot, however, be documented. Hence any faults that cannot be detected by means of the device documentation alone can be recognised at an early stage.

FORUM

Ch. Wolf:
Validation of Heat Sealing Equipment

LETTERS

Dicussion on BD-Test

EDUCATION

P. Heeg
Echinococcosis

MAILBOX

Documentation with a datalogger system

RECOMMENDATIONS BY THE QUALITY TASK GROUP (AK »QUALITÄT«)

Bowie Dick Test


Publisher
Information on this page provided with kind permission of: mhp-verlag GmbH
Address: Marktplatz 13
65183 Wiesbaden
Germany
Web: www.mhp-verlag.de
Email:
mhp stands for publications in medicine, hygiene and prevention. For more than 20 years, the mhp publishing house has devoted itself to supporting developments aiming for a better standard of hygiene and infection control in hospitals and other medical facilities.
Archive
WFHSS - World Forum for Hospital Sterile Supply:: Publications :: Periodicals :: Central Service :: Central Service Archive :: Central Service 5/01

© 2008 WFHSS • Updated: 29 October 2006, 04:09 [GMT]