WFHSS - World Forum for Hospital Sterile Supply

Mitglieder [17.11.2008] new WFHSS member organization: IHCC - Iranian Health Care centers (Iran)
Kommunikation [20.11.2008] Open letter to heads of sterilization departments in countries where there are no national societies
Konferenzen & Kongresse [19.11.2008] current Conference: Annual NZSSA Conference 2008 "Join the Rush - Knowledge is Gold" - 19/20/21 November 2008 - Queensland, New Zealand
Kurse & Seminare
Fragen & Antworten [20.11.2008] answer to question 1989: Surgical instruments magnetized
Diskussionsforum
CEN
Publikationen [14.11.2008] new publication: Central Service Issue 5/08
Weiterbildung [17.11.2008] Transmissible spongiform encephalopathy agents: safe working and the prevention of infection Guidance from the Advisory Committee on Dangerous Pathogens' TSE Working Group by United Kingdom Department of Health - ACDP TSE Working Group
WFHSS-WebShop
Industrielle Partner [18.11.2008] new WFHSS Industrial Partner: ArjoWiggins - Medical & Hospital Paper
Commercial Announcements
Related Sites & Links [18.11.2008] link to new Industrial Partner ArjoWiggins added
Stellenangebote
Humor
Industrielle Partner:

8219901

Entwickelt von Baumeister Mediasoft Engineering

WFHSS - World Forum for Hospital Sterile Supply

Central Service - Issue 2/02 - English VersionZentralsterilisation - Ausgabe 2/02 - Deutsche Version

StartseiteLogosAktuellesSucheSeitenübersichtKontakt

WFHSS - World Forum for Hospital Sterile Supply:: Publikationen :: Fachzeitschriften :: Zentralsterilisation :: Zentralsterilisation - Archiv :: Zentralsterilisation 2/02
Zentralsterilisation - Ausgabe 2/02

Zentralsterilisation - Ausgabe 2/02

Die folgende Information ist im Augenblick leider nicht in Deutsch verfügbar. Wir bitten um Ihr Verständnis.

EDITORIAL

NEWS UPDATE

What's New in Standardisation: Sterile Supply, Air Pockets and Dataloggers
In memoriam: Jack van Asten
Prions: Challenges in Disinfection and Sterilisation
Forum 2002: Medical Device Processing - Verification of Performance Parameters
Hospital Cleaning and Surface Disinfection. 4th Kassel Symposium on Hospital Hygiene

MAIN ARTICLES

M.C. Gaspar, B. Peláez, C. Fernández, J. Fereres:
Microbiological Efficacy of Sterrad 100S and LTSF Sterilisation Systems Compared to Ethylene Oxide
(Zentr Steril 2002; 10 (2): 91-99)

Abstract
Objective: To compare the efficacy of the Sterrad 100S (S100S) and low temperature steam formaldehyde sterilisation, 50 and 60 °C cycles (LTSF50 and LTSF60), to the efficacy of the ethylene oxide system (EtO).

Material and methods: Smooth and rough surfaced carriers were inoculated with approximately 107 gram-positive or gram-negative vegetative organisms, mycobacteria, B. subtilis spores or C. albicans, under clean or dirty (10 % blood or 0,9 % NaCl) conditions. These carriers were then subjected to different sterilisation methods. The residual bacterial load after sterilisation was determined. Counts of unprocessed positive controls were determined and log10 reductions were calculated. Interactions between sterilisation methods and type of suspension in comparison with EtO under clean conditions were assessed using a multivariate regression lineal model (GLM).

Results: Both LTSF cycles were more effective than EtO. This is due to the fact that LTSF efficacy was unaffected by the presence of salt. The coefficients and 95 % CI obtained in the GLM were: LTSF50 : 2.22 (95 % CI 0.41; 4.03), LTSF60 : 2.25 (95 % CI 0.44; 4.06) and S100S: -1.16 (95 % CI -2.97; 0.64). Under clean conditions and in presence of 10 % blood, all methods achieved a mean reduction higher than 6 log10, and there was no difference between methods. Under salty conditions there was a statistical difference between methods (p < 0.001). B. subtilis spores and gram-positive bacteria were the most resistant microorganisms to all sterilisation methods (p < 0.001).

Conclusions: LTSF (both cycles) was the most efficacious method, and was the only method unaffected by the presence of salt. In the absence of salt all methods showed similar efficacy.

Working Group Instrument Preparation:
Sterilisability of Reusable Surgical Instruments
Part 2: Study Results - Implications for Working Practices in the CSSD
(Zentr Steril 2002; 10 (2): 100-109)

Abstract
Quality assurance is being ascribed increasingly more importance in the hospital setting. In the circuit traversed by medical devices undergoing sterilisation, quality assurance is intended as a means of guaranteeing the user in the hospital that impeccable devices are being used, while at the same time giving the patient the feeling of high-quality care. The sterilisation study presented here furnishes proof that surgical instruments can be sterilised with steam and thus meet the essential requirements set out in Annex 1 of the Medical Devices Directive 93/42 EEC. Part l of this publication described how to design process challenge devices for steam sterilisation. This second part of the publication presents the results of the study and elaborates on their implications for routine working practices in the CSSD.

EDUCATION

P. Heeg
Disease Caused by Lice (Pediculosis)

MAILBOX

Space Requirements

RECOMMENDATIONS BY THE QUALITY TASK GROUP (AK »QUALITÄT«)

Why Computerised Documentation? - Reflections from the Field


Herausgeber
Informationen auf dieser Seite mit freundlicher Genehmigung von: mhp-verlag GmbH
Adresse: Marktplatz 13
65183 Wiesbaden
Germany
Web: www.mhp-verlag.de
Email:
mhp steht für Medizin, Hygiene und Prävention. Der mhp-Verlag ist seit mehr als 20 Jahren Spezialist für Publikationen im Bereich der Hygiene.
Archiv
WFHSS - World Forum for Hospital Sterile Supply:: Publikationen :: Fachzeitschriften :: Zentralsterilisation :: Zentralsterilisation - Archiv :: Zentralsterilisation 2/02

© 2008 WFHSS • Aktualisiert: 29. Oktober 2006, 04:09 [GMT]