WFHSS - World Forum for Hospital Sterile Supply

Mitglieder [17.11.2008] new WFHSS member organization: IHCC - Iranian Health Care centers (Iran)
Kommunikation [20.11.2008] Open letter to heads of sterilization departments in countries where there are no national societies
Konferenzen & Kongresse [19.11.2008] current Conference: Annual NZSSA Conference 2008 "Join the Rush - Knowledge is Gold" - 19/20/21 November 2008 - Queensland, New Zealand
Kurse & Seminare
Fragen & Antworten [20.11.2008] answer to question 1989: Surgical instruments magnetized
Diskussionsforum
CEN
Publikationen [14.11.2008] new publication: Central Service Issue 5/08
Weiterbildung [17.11.2008] Transmissible spongiform encephalopathy agents: safe working and the prevention of infection Guidance from the Advisory Committee on Dangerous Pathogens' TSE Working Group by United Kingdom Department of Health - ACDP TSE Working Group
WFHSS-WebShop
Industrielle Partner [18.11.2008] new WFHSS Industrial Partner: ArjoWiggins - Medical & Hospital Paper
Commercial Announcements
Related Sites & Links [18.11.2008] link to new Industrial Partner ArjoWiggins added
Stellenangebote
Humor
Industrielle Partner:

8219763

Entwickelt von Baumeister Mediasoft Engineering

WFHSS - World Forum for Hospital Sterile Supply

Central Service - Issue 6/02 - English VersionZentralsterilisation - Ausgabe 6/02 - Deutsche Version

StartseiteLogosAktuellesSucheSeitenübersichtKontakt

WFHSS - World Forum for Hospital Sterile Supply:: Publikationen :: Fachzeitschriften :: Zentralsterilisation :: Zentralsterilisation - Archiv :: Zentralsterilisation 6/02
Zentralsterilisation - Ausgabe 6/02

Zentralsterilisation - Ausgabe 6/02

Die folgende Information ist im Augenblick leider nicht in Deutsch verfügbar. Wir bitten um Ihr Verständnis.

EDITORIAL

Norms and Directives in the Light of 10 New Countries accessing the EU

NEWS UPDATE

What's New in Standardisation: Large Steam Sterilizers
Regulations on Reuse of Single-Use Medical Devices to be Reviewed in 2004
Infection Risks Posed by Returned Medical Devices: BVMed Recommends an Approach
3rd Schleswig Holstein CSSD Forum on 26 October in Ostseebad Dam

MAIN ARTICLES

A.W. van Drongelen*, A.C.P. de Bruijn, J.F.M.M. Huys, B. Muis:
Optimisation of the Process for a Manually-Operated Jacketed Steam Steriliser
(Zentr Steril 2002; 10 (6): 373-384)

Abstract
The objective of this study was to improve the sterilisation process for a manually-operated jacketed steam steriliser of a type that is mainly used in developing countries. The experiments were focussed on steam penetration and drying for both a standard textile test pack and a full load of three sterilisation drums filled with textiles, while keeping the power and water consumption as low as possible. The process should fulfil the requirements for steam penetration and drying specified in European standards.

Hardly any steam penetration was observed in a standard textile pack during the sterilisation process as recommended by the manufacturer. The process could not be considered sterilisation. For the improvement of the steam penetration, most attention was paid to steam flushing and steam pulsing.

A full load of textiles in three sterilisation drums proved to be the most difficult to sterilise textile load. This load could be sterilised with the use of a modified sterilisation cycle. This cycle consisted of 20 minutes steam flushing at atmospheric pressure, three times a slow pressure build-up to 300 kPa followed by a pressure reduction to 100 kPa and finally a slow pressure build-up to the sterilisation pressure and sterilisation for four minutes.

The vacuum created with the standard steam ejector proved to be insufficient to dry the load and the use of water ring pumps was considered to be inappropriate. An external condensation vessel was developed. This vessel proved to be a low cost means to create an effective drying vacuum.

M. Pfeifer:
Validation of SonoCheck for the Monitoring of Ultrasonic Energy of Ultrasonic Cleaners
(Zentr Steril 2002; 10 (6): 386-391)

Abstract
SonoCheck, a new test for monitoring the ultrasonic energy in ultrasonic cleaner used for reprocessing surgical instruments was successfuly validated. In the validation tests it was shown that SonoCheck will indicate ultrasonic energy needed to induce cavitation and give additional informations regarding the energy level. Thermal disinfection will not influence the result of SonoCheck. Due to the design of the product no residue will be left in the bath when performing the test and therefore SonoCheck can also be used to check the ultrasonic energy during the actual reprocessing of surgical instruments for load safety.

PRACTICAL TIPS

Sterilisation Section of the German Society for Hospital Hygiene (DGKH): The Problems Arising from Non-Condensable Gases (NCGs) in Steam Sterilisation Processes with a Pulsed Vacuum Procedure/Instructions for Interpreting the Inert Gas Value of 3.5% in DIN EN 285

RECOMMENDATIONS

P. Hooper
Choosing the Right Cycle

DISCUSSION

W. Renders:
Can chemical indicators replace biological ones?

U. Kaiser:
Which Sterilant is the Active Component in a Steam Sterilisation Process? Steam? No, it's water!

N. Ghassemieh:
Reprocessing Medical Devices - New Regulations for Ensuring Safety and Economic Efficiency

MAILBOX

Textile Filters/Demonstration of Cleaning Results

RECOMMENDATIONS BY THE QUALITY TASK GROUP (AK »QUALITÄT«)

Water for Operating Washer-Disinfectors


Herausgeber
Informationen auf dieser Seite mit freundlicher Genehmigung von: mhp-verlag GmbH
Adresse: Marktplatz 13
65183 Wiesbaden
Germany
Web: www.mhp-verlag.de
Email:
mhp steht für Medizin, Hygiene und Prävention. Der mhp-Verlag ist seit mehr als 20 Jahren Spezialist für Publikationen im Bereich der Hygiene.
Archiv
WFHSS - World Forum for Hospital Sterile Supply:: Publikationen :: Fachzeitschriften :: Zentralsterilisation :: Zentralsterilisation - Archiv :: Zentralsterilisation 6/02

© 2008 WFHSS • Aktualisiert: 29. Oktober 2006, 04:09 [GMT]